Model 532.53+8xx Absolute Pressure Gauge with Switch Contact
Brand: WIKA Instruments Ltd.Wherever the process pressure has to be indicated locally and, at the same time, circuits need to be switched, the model 532.5x switchGAUGE finds its use.
Features- High overload safety, long service life due to metal media chamber sealing
- Media chamber protected against unauthorised access
- Also available with liquid-filled case for high dynamic pressure loads or vibrations
- Instruments with inductive contacts for use in hazardous areas
- Instruments with switch contact for PLC applications
Specifications
Design
- Mechanical absolute pressure gauge per DIN 16002
Nominal Size in mm
- 100
- 160
Accuracy Class
- 1.6 (model 532.53)
- Option:
- 1.0 (model 532.52)
- 2.5 (model 532.54)
- The measurement accuracy is ensured for ambient pressure fluctuations between 955 and 1,065 mbar (min. and max. of atmospheric pressure)
Scale Ranges
- 0 to 25 mbar 2) to 0 to 25 bar Absolute Pressure [0 to 0.36 psi to 0 to 3,600 psi Absolute Pressure]
Scale
- Single scale
- Option: Dual Scale
Pressure Limitation
- Steady: Full scale value
- Fluctuating: 0.9 x full scale value
Overload Safety
- 10 x full scale value, max. 25 bar absolute pressure, min. 1 bar absolute pressure
- Option: 20 x Full Scale Value, max. 25 bar Absolute Pressure, min. 1 bar Absolute Pressure
Process Connection with Lower Measuring Flange
- Small flange for vacuum applications DN 10
- Small flange for vacuum applications DN 16
- Open connecting flange DN 25 PN 25, DIN 2501, form D per DIN 2526
- Open connecting flange DN 25 PN 25 per EN 1092-1, form B
- 1/2" NPT Female
- 1/2" NPT
- G ½ B
Permissible Temperature
- Medium: +100° C [+212° F] maximum
- Option: +200° C [+392° F] maximum
- Ambient: -20 ... +60° C [-4 ... 140° F]
- Option: -40 ... +60° C (silicone oil filling) ¹⁾
Temperature Effect
- When the temperature of the measuring system deviates from the reference temperature (+20° C): max. ±0.8 %/10 K of Full Scale Value
Case
- Option:,Safety version S3 per EN 837: With solid baffle wall (Solidfront) and blow-out back (models 532.32
- 532.33, 532.34)
- Instruments with liquid filling with compensating valve to vent case
- Version S1 per EN 837: With blow-out device in case back
Case Filling
- Without
- Option: With silicone oil M50 case filling, ingress protection IP65 (models 533.52, 533.53, 533.54)
Wetted Materials
- Diaphragm element (pressure element), ≤ 0.25 bar: Stainless Steel 316Ti, > 0.25 bar: NiCr Alloy (Inconel)
- Media chamber with process connection: Stainless Steel 316L
- Option: Wetted parts from Monel (models 56x.3x, 56x.5x)